Product Description:
Flash Point is defined as “the lowest temperature of a liquid at which its vapours will form a combustible mixture with air”. It is a convenient and reliable classification of the “flammability” of many substances.
- Open Cup tests simulate an un-contained condition, for example a spillage.
- Closed Cup tests simulate an enclosed environment, for example storage in a tank or sealed container and are usually specified due to improved precision.
Most industries need to test the Flash Point of raw materials, products or waste to ensure:
Product quality - as a measure of consistency and performance comparison
Compliance testing - to test safety classification for handling, storage, transport and waste
In-service analysis - tests on in-use oils and other substances for contamination / adulteration
General safety - to evaluate hazard potential
Specifications - to check conformance
A wide range of features ensures ease of operation and requires minimum operator skill. Test parameters are set via the keypad and a digital display provides temperature, sample size, test time and flash detection status. Default ‘Auto’ values are available for standard test conditions.
The Elcometer 6910 Closed Cup Tester can test from ambient to 300°C (572°F). Temperature is factory calibrated but facilities for user verification and calibration are incorporated.
Flash Point is automatically detected using a thermally activated detector, reducing the risk of operator error and minimising the potential danger of inhaling fumes during a test. A rechargeable gas tank with On/Off switch and fine adjustment are integral to the unit. Supplied with a 2ml (0.067fl oz) syringe and ignitor.
To test below the ambient temperature, an optional cooling module, suitable for temperatures as low as 0°C (32°F), should be ordered.
Technical Specification:
Part Number |
Description |
UK 240V |
EUR 220V |
US 110V |
K0UK6910M010 |
K0006910M010 |
K0US6910M010 |
Elcometer 6910/1 Setaflash Series 3 Closed Cup Tester |
Sample Size and Temperature Range |
2ml (0.067fl oz) for Flash Points, from ambient up to 100°C (212°F) |
4ml (0.135fl oz) for Flash Points between 100°C to 300°C (212°F to 572°F) |
Test Times |
Between 1 and 30 minutes (user definable) |
Default Values |
1 minute for Flash Points, from ambient up to 100°C (212°F) |
2 minutes for Flash Points between 100°C to 300°C (212°F to 572°F) |
Cup Material |
Aluminium |
Dimensions |
256 x 280 x 256mm (10.1 x 11.0 x 10.8”) |
Weight |
4kg (8.8lb) |
Certificate available |
Certificate of Calibration: issued for Fixed Calibration equipment and shows readings and traceability. |
Packing List |
Elcometer 6910 Setaflash ‘Series 3’ closed cup tester, 2ml (0.067fl oz) syringe, gas canister and ignitor, silicone rubber tubing for gas canister to gas jet connection, silicone sample well ‘O’ ring seal for cup/lid red coloured), viton sample well ‘O’ ring seal for cup/lid (black coloured), mains power cable, automatic flash detector probe, Series 3 manuals (on CD) and operating instructions |
Can be used in accordance with:
ASTM D 1655, ASTM D 3278, ASTM D 3828, ASTM D 3934, ASTM E 502, BS 3900-A13, BS 3900-A11 BS 3900-A14,
BS 6664-3, BS 6664-4, EN 456, ISO 3679, ISO 3680
Standards in grey have been superseded but are still recognised in some industries.
Accessories:
KT006910N001 |
Cooling Module Option for 0°C - Ambient Temperature (32°F - Ambient Temperature) |
|