Download Full Catalog
Product Description:
The Elcometer 6075/1 SP60 is an affordable sphere spectrophotometer, designed to give fast, precise and accurate colour measurement information on materials ranging from paper and paint to plastics and textiles.
- Lightweight, compact, portable instrument
- Diffuse/8° sphere optical geometry
- Fixed 8mm aperture
- Large, easy-to-read graphical LCD display
- Opacity and colour strength measurement
- Simultaneous measurement of both specular component included and specular component excluded
- Rugged construction
- Rechargeable battery for remote use
Key Features
- Measuring Functions and Indices
The Elcometer SP60 provides absolute and difference measurements for the following colourmetric systems. These values can be obtained from any of the nine illuminants with 2° or 10° observer angle: L*a*b*, DL*Da*Db*, L *C*h°, DL*DC*DH*, DE*ab, DECMC, DE CIE94 and XYZ. Whiteness and Yellowness per ASTM E 313-98.
- Pass/Fail Mode
The instrument stores up to 1024 standards with tolerances for easy pass/fail measurement. A red/green LED indicator and the LCD display provide visual confirmation of results. A tone also sounds to indicate a fail result and measurement completion.
- Quick Colour Compare
A quick measurement can be taken to compare two colours. This allows the operator to take quality control readings in a time efficient manner without having to create tolerances or store data.
- The Sphere
The Elcometer SP60’s diffusing sphere is made of Spectalon®, a durable, highly reflective material designed to perform in a rigorous production environment. The diffusing material prevents degradation due to the flaking and chipping of the sphere wall material.
- Opacity, Colour Strength and Shade Sorting
The instrument can measure opacity as well as three colour-strength options: chromatic, apparent and tri-stimulus calculations. The Elcometer SP60 also performs 555 shade sorting. This is an important consideration in the colour quality control of manufactured products involving plastics, painted or textile materials.
- Texture and Gloss influence
To determine the influence of the specular component, the SP60 allows simultaneous measurement of both specular - included (colour) and specular-excluded (appearance)
- User friendly Ergonomics
In addition to on-board programmes to assist the operator in the measurement process, the instrument itself is highly user -friendly. It is compact and lightweight with a wrist strap and the tactile side grips make it easy to hold. Read-outs are large and easy to see. A rechargeable battery pack allows extended operation of the instrument.
Technical Specification:
Part Number |
Description |
UK 240V |
EUR 220V |
US 110V |
K0UK6075M001 |
K0006075M001 |
K0US6075M001 |
SP60 Portable Sphere Spectrophotometer 8mm Fixed Aperture |
Measuring Geometrics |
d/8°, DRS spectral engine, fixed aperture: 8mm viewing/12mm illumination |
Light Source |
Gas filled tungsten lamp |
Illuminant Types |
C, D50, D65, D75, A, F2, F7, F11, F12 |
Standard Observers |
2° and 10° |
Spectral Range |
400-700nm |
Memory |
1,024 standards with tolerances, 2,000 samples |
Measurement Range |
0 to 200% reflectance |
Measuring Time |
Approximately 2 seconds |
Inter-Instrument Agreement |
CIE L*a*b*:
Average 0.40 ΔE*ab based on average of 12 BCRA Series II tiles
(specular component included)
Maximum 0.60 ΔE*ab on any tile (specular component included)
CMC Equivalent:
Average 0.30 ΔEcmc based on average of 12 BCRA Series II tiles
(specular component included)
Maximum 0.50 ΔEcmc on any tile (specular component included) |
Short-term Repeatability* |
0.10 ΔE*ab on white ceramic (standard deviation) |
Lamp Life |
Approximately 500,000 measurements |
Power Supply |
Removable (Ni-metal hydride) battery pack; 7.2DC rated @1450mAh |
Measurements per Charge |
1000 measurements within 8 hour period |
Weight |
1.1kg (2.4lb) |
Dimensions |
109 x 83 x 195mm (4.3 x 3.3 x 7.7”) |
Certificate Available |
Certificate of Calibration: issued for Fixed Calibration equipment and shows readings and traceability. |
Packing List |
Elcometer 6075/1, calibration standards, AC adaptor, carry case & operating instructions |
Only available for sales within the EU
*Based on 20 measurements on a white tile
Can be used in accordance with:
AS/NZS 1580.601.3, ASTM C 609, ASTM D 2244, ASTM E 1164, ASTM E 308, ASTM E 313, BS 8493, DIN 5033-2,
DIN 5033-3, DIN 5033-4, DIN 5033-7, DIN 6174, EN 12373-12, EN 13523-15, ISO 7724-2, ISO 7724-3, JIS K 5600-4-5, JIS K 5600-4-6, NF T36-006,
NF X08-012-1, NF X08-012-2
Accessories:
UK 240V |
EUR 220V |
US 110V |
Description |
KTUK6075P001 |
KT006075P001 |
KTUS6075P001 |
Battery Charger Kit |
KT006075P002 |
KT006075P002 |
KT006075P002 |
NiMH Battery Pack |
Download Full Catalog |