Back to S
Series Products
Atomic
Absorption Spectrophotometer, Double
Beam Atomic Absorption Spectrophotometer, Single
Beam Atomic Absorption Spectrophotometer, NIR
Spectrophotometer, Spectrofluorometer,
Bio
- Spectrophotometer, Double
Beam UV-VIS spectrophotometer, Single Beam UV-VIS Spectrophotometer,
UV-VIS
Spectrophotometer, Scanning
Visible Spectrophotometer, Visible
Spectrophotometer, Fluorometer,
Colorimeter
Spectroscopy - Single Beam UV-VIS Spectrophotometer
Features
|
Scanning
Single beam UV-VIS Spectrophotometer with programmable scan
speeds |
Microprocessor based with Printer Interface |
Optional
PC compatibility with RS 232 C Interface and MS Windows® based
software for data acquisition, processing, storage, retrieval
& interpretation of data |
Fully automatic,
user friendly and menu driven |
Automatic
wavelength calibration, selection of wavelength and positioning
of sample holder |
Compact,
modular & ergonomically designed |
Stable single
beam optics |
Automatic
6 position cuvette holder for 10 mm pathlength cuvettes |
Lamp selection
to conserve life of the light sources |
Programmable wavelength for lamp change over |
Advanced Data Processing Functions
Stand Alone
- Scanning
- Discrete Wavelength
- Concentration by standards and
- factor methods
- Time scan with minimum time interval of 0.1 sec.
- Self check and self diagnostic
Extra with PC Option
- Upto 4th order Derivate Spectra
- Overlay of maximum 10 Spectra
- Ratiometric user programmable equations
- Peak Picking
- Spectra Zooming
- Data saving & retrieving
Description
This SINGLE BEAM UV-VIS SPECTROPHOTOMETER SL 159 is a rugged, cost
effective, microprocessor version of spectrophotometer introduced
to fill in the need for routine analysis where variation of parameters
is very critical.
Specifications
Spectral |
Range |
200 to 1000 nm |
Bandwidth |
2 nm |
Accuracy |
± 0.5 nm |
Readability |
0.1 nm |
Repeatability |
± 0.2 nm |
Drive Speed |
300, 600, 900 nm/min |
|
Photometric |
Range |
± 2.5 Abs |
Accuracy |
± 0.005 Abs. at 1.0 Abs.
± 0.010 Abs. at 1.5 Abs. |
Repeatability |
± 0.002 Abs. at 1.0 Abs. |
Stability (Baseline) |
± 0.003 Abs/hr.
after 2hour warm up |
Stray light |
< 0.1% at 220 nm
with Nal 10 g/L |
|
Light Source |
Deuterium (D2) &
Tungsten(W) Halogen Lamps |
|
Monochromator |
Czerny-Turner type with 1200 lines/mm holographic
grating |
|
Detector |
Wide range Photo Diode |
|
Control |
Microprocessor (Computer - Optional) |
|
Data
Processing |
Microprocessor Mode |
Spectrum, Discrete ,
Concentration and Time Scan |
PC Mode |
Spectrum, Overlay, Time scan, Discrete ,
Ratiometric, Concentration, -factor,
Upto 4th Order Derivate, Peak picking, spectra zooming and Data
saving & retrieving |
|
Data
Presentation |
Display |
24 x 2 Dot Matrix LCD Display |
|
Readability |
%T |
0.1 |
Abs |
0.001 |
Conc |
± 1 Count |
Hard copy |
On printer |
|
Power Requirement |
230 V ± 10%, 50 Hz, 1 , Max. 150 VA |
|
Physical Dimensions |
Size (WxDxH) in mm |
570 x 485 x 145 |
Weight in kg |
Approx. 22 |
|
Accessories supplied |
With instrument |
10 mm pathlength cuvettes |
At Extra cost |
PC, UV-VIS Spectrasoft Software and Printer |
Specifications subject to change due to continuous development
|
|
|