headerphoto
Laboratory Equipments >> W Series >> Waste Sterilizer
High Pessure High Speed Horizontal Cyliderical / Rectangular Regulated Medical Waste Sterilizer
Waste Sterilizer

Waste Sterilizer

Purpose:- Equipment is designed for sterilization of regulated biomedical waste generated from hospital for destruction of bacterial and related organism (Spores).

Design:- The sterilizer equipment to treat medical waste is designed for following parameters.

1. For gravity flow autoclave

The temperature of not less than 121ºC and pressure of not less than 1.02 K.G./CM2 and autoclave residence time not less than 60 minutes.
The temperature of not less 135ºC and pressure of not less than 2.1 K.G./CM2 and autoclave residence time not lesss than 45minutes.
The temperature of not less 149ºC and pressure of not less than 3.5 K.G./CM2 and autoclave residence time not lesss than 30 minutes.

2. For vacuum autoclaving with pre vac & purge system.

The temperature of not less than 121ºC and pressure of not less than 1.02 K.G./CM2 and autoclave residence time not less than 45 minutes with one or more pre vacuum pulse.
The temperature of not less than 135ºC and pressure of not less than 2.1 K.G./CM2 and autoclave residence time not less than 30 minutes with one or more pre vacuum pulse.

Construction:- The equipment is constructed as per the ASME section VIII.

Shape:- Either cylindrical/rectangular. Chamber is made off boilder grade sheet/AISI 304/AISI 316 thick sheet having one or two nos. pivoted door mounted on a sturdy structural stand dully insulated & covered either by S.S. of M.S. Sheet. Pit mounted desiggn is also available as per the customer requirement for easy loading & unloading. The door will be provided with automatic safety door looking / unlocking mechanism. Valves are provided for operation safety steam inlet, drainage etc.

Control and Monitoring System:- Microprocessor based computer system and printer is provided for controlling and monitoring the sterilizing cycle with the following feature:

Fully programmable microprocessor based unit with no volatile memory and with different program combinations.

Automatic operation of the sterilizer.
Self diagnostic system gives fault massage to the operator.
Manual operation of the sterilizer
Self diagnostic system gives fault massage to the operator.

Cycle progress and information indicating any error massage is displayed and recorded on the printer. The printer shall also record data time, day load identification no. and operating parameters such as temperature of the steam in chamber, steam pressure, vacuum time and temperature of steam condensate after cooling through out the entire length of autoclave cycle.

Available in following Sizes:-

Cylindrical Model:-
400 (Dia) x 600 mm (Depth)
500 (Dia) x 1200 mm (Depth)

Rectangular Model:-
600 (H) x 600 (W) x 1200 mm (D)
900 (H) x 600 (W) x 1500 mm (D)

Other custom built sizes are also available.